Chemiluminescence Detection System for Nitrosamines

FANG Jie,JI Min,SUN Miao,LI Fei,ZHU Zhen,LIU Yong,WANG An
DOI: https://doi.org/10.3969/j.issn.1004-8138.2011.04.004
2011-01-01
Abstract:A chemiluminescence devices for the nitrosamines gas chromatography was build,that can detected the concentration of the NO from the gas chromatography with devices as the gas chromatography detector,according to the concentration of the nitrosamines was in proportional to the NO concentration,then concentration of nitrosamines can be calculated.The some key design and how affected our system performance by the chamber pressure were investigated,and the best experimental parameters were confirmed.The NO standard gas with concentration of 4.2mg/kg was practically determined by the detection system,and linearity for the system was more than 0.98,and detection limitation was less than 1pg/s.The system can quickly and accurately detect trace NO,and the NO can be the chromatography detector for nitrosamines detect.