Determination of Nitrobenzenes in Air by a Needle Trap Device (NTD) and Gas Chromatography-Mass Spectrometry (GC-MS)

Xian Dong Qian Jin Haifeng Zou Yong Li Wanli Wei Zhuo Chen a School of Chemistry and Materials Science,Guizhou Normal University,Guiyang,Chinab No.2 High School,Guiyang,Chinac School of Chemical Engineering,Guizhou MinZu University,Guiyang,China
DOI: https://doi.org/10.1080/00032719.2024.2418861
2024-10-29
Analytical Letters
Abstract:A simple needle-trap device (NTD) coupled with gas chromatography-mass spectrometry (GC-MS) was developed to determine trace levels of nitrobenzenes (NBs) in workplace air. The NTD with sorbent inside the needle was employed for the quantification of nitrobenzene, o-nitrotoluene, m-nitrotoluene, p-nitrotoluene, o-nitrochlorobenzene, m-nitrochlorobenzene, and p-nitrochlorobenzene. The optimum temperature and time for desorption of analytes were 270 °C and 4 min, and parameters such as sampling flow rate, breakthrough volume, and storage capability were also optimized. The limits of detection and quantification were in the ranges of 1.8 × 10 −4 –1.1 × 10 −3 and 5.9 × 10 −4 –3.6 × 10 −3 mg m −3 , respectively. The developed protocol was successfully applied for the determination of NBs in vehicle repair plants where the highest detected concentration was 0.11 mg m −3 . The established NTD-GC-MS generally applies to monitoring trace levels of NBs in the air.
chemistry, analytical