Determination of huperzine A in huperzine A sustained-release patches by high performance liquid chromatography(HPLC)

WU Ting,ZHANG Jun,CHU Ji-hong,JU Wen-zheng,XU Mei-juan,CHEN Min,TAN Heng-shan
DOI: https://doi.org/10.3969/j.issn.1671-2560.2011.09.004
2011-01-01
Abstract:Objective To determine the content of huperzine A in huperzine A sustained-release patches.Methods The huperzine A sustained-release patches were detected by high performance liquid chromatography(HPLC) on Agilent Zorbax Ehipse DXB-C18 column with 0.8% triethylamine(pH 10.5,V/V 55:45) as the mobile phase,a detection wavelength of 313 nm,a flow rate of 1.0 mL/min,a column temperature of 30 ℃ and a sample amount of 20 μL.Results The huperzine A showed a favorable linearity in a range from 0.155 to 2.472 μg/ml(r=0.9996).The mean recovery was 96.44%(RSD=1.14%).Conclusion The method in this study appeared simple and exclusively accurate,which justifies its use in quality control of huperzine A preparations.