On-line distillation and solid phase extraction preconcentration of volatile phenolic compounds in water samples with detection by spectrophotometry

CHEN MingLi,HU HaiZhu,WANG JianHua,WANG Jing,HUO YuMei
DOI: https://doi.org/10.1360/032010-550
2011-01-01
Abstract:A novel sequential injection on-line distillation procedure for volatile phenolic compounds is developed. The derivative between phenol and 4-aminoantipyrine (4-APP) is adsorbed onto an Amberlite XAD-7 resin packed mini-column. The phenol-4-APP derivative retained on the mini-column is effectively recovered by using a small amount of ethanol followed by spectromphotometric detection. The present procedure is suitable for the direct determination of volatile phenolic compounds at mg/L levels. It can also be used for the quantification of much lower concentration of volatile phenolic compounds at μg/L level after preconcentration with a larger sample volume. With the sample loading volume of 4 mL, a linear range of 0.004–0.3 mg/L is obtained along with a detection limit is 0.002 mg/L. A relative standard deviation of 1.4% (n = 9) is achieved at 0.1 mg/L. The procedure is applied to the determination of low level of volatile phenolic compounds in water samples and spiking recoveries within 96%–102% are achieved.
What problem does this paper attempt to address?