TI Value of Flow Sensitive Alternating Inversion Recovery Effect on MR Perfusion Imaging of Pulmonary Parenchyma

FAN Li,LIU Shi-yuan,XIAO Xiang-sheng,JIANG Qing-jun
DOI: https://doi.org/10.3969/j.issn.1002-1671.2006.08.010
2006-01-01
Abstract:Objective To find out the optimal TI value of FAIR in order to acquire high quality of MR perfusion imaging of lung.Methods 12 adults without thoracic and cardiovascular diseases volunteered participation in this study. All cases underwent FAIR imaging with different TI values on a 1.5T whole body scanner (GE medical system). TI values were 800 ms,1000 ms,1200 ms,1400 ms and 1600 ms,respectively.The changes of signal intensity between control and tag images in lung and aorta were evaluated, signal to noise ratio (SNR)of perfusion images was also evaluated with different TI values.Results ①Signal intensity change of lung: no statistic difference between 800 ms and 1600 ms (P>0.05), there were significant statistic differences of any other two TI values (P<0.05),the change was maximum at 1000 ms;②Signal intensity change of aorta: no statistic difference between 1400 ms and 1600 ms (P>0.05), there were significant statistic differences of any other two TI values (P<0.05),the change was maximum at 1000 ms;③SNR of perfusion images: no statistic differences among 1000 ms , 1200 ms , 1400 ms and 1600 ms ( P>0.05 ) , there were significant statistic differences of any other two TI values (P<0.05).Conclusion Signal intensity change between control and tag images in lung and aorta is maximum at 1000 ms, and the SNR is relatively good. So TI value of 1000 ms is recommended.