Renal Perfusion Imaging Using FAIR and FIESTAat 3.0T MR Scanner

YANG Xue-dong,WANG Jing,LIU Dong-dong,DANG Yi,WANG Xiao-ying,ZHANG Jue,FANG Jing
DOI: https://doi.org/10.13929/j.1003-3289.2011.08.009
2011-01-01
Abstract:Objective To develop a renal arterial spin labeling(ASL) using flow-sensitive alternation inversion recovery(FAIR) and fast imaging employing steady-state acquisition(FIESTA) on a 3.0T MR scanner.Methods FAIR prepared single-shot FIESTA was used.Both breathhold and non breathhold were used to acquire ASL image.Five healthy subjects were included in this study.The feasibility and quality of images were assessed,and renal blood flow was calculated.Results The scan time was 20 s for breathhold method,and 55 s for non breathhold method.ASL images were successfully acquired for all 5 subjects.The quality of ASL images for both methods was acceptable.The renal blood flow(RBF) for breathhold were(220±41)ml/(100 g·min) in the whole kidney,(303±65)ml/(100 g·min) in the cortex,(103±11) ml/(100 g·min) in the medulla,while for nonbreathhold were(221±52) ml/(100 g·min) in the whole kidney,(302±67)ml/(100 g·min) in the cortex,and(107±17)ml/(100 g·min) in the medulla.Conclusion FAIR-FIESTA can be used to acquire renal ASL images,and breathhold scanning is recommended.