Determination of Dichlorvos Residues by Flow-injection Chemiluminescence Method

BAO Chang-li,FAN Xi-yue,SONG Hai-ou,XU Ren-shun,ZHANG Xiao-yu
DOI: https://doi.org/10.3321/j.issn:1671-5489.2006.04.028
2006-01-01
Abstract:A new flow injection chemiluminescence(FI-CL) method for the determination of dichlorvos was developed.The method is based on the CL reaction of dichlorvos with luminol-(H_2O_2) in the alkaline medium and sensitized by the nonionic surface-active agent polyoxyethyle octyl phenyl ether.The results show that the CL intensity is linear to the concentration in the range of(8.00×10~(-8))~(2.50×10~(-5)) g/mL,with a detection limit of(2.20×)(10~(-8)) g/mL.The standard addition recovery and relative standard deviation(n=6) of three samples were(94.0%~)104% and(4.27%~5.75%,) respectively.This method has been applied to the(determination) of dichlorvos residues in crops and vegetables.