A Contrast Study of the Performance of Carotid Surface Coil and Volume Coil in 3.0T Neurovascular MRI

朱高杰,李斌,王杰
DOI: https://doi.org/10.3969/j.issn.1674-1633.2009.05.006
2009-01-01
Abstract:Objective To compare the performance of 8-channel phased array carotid surface coil over a conventional 16-chanel volume coil,the signal-to-noise ratio(SNR) and image uniformity in 3.0 Tesla magnetic resonance imaging system had been studied.Materials and Methods A phantom filled with mineral oil from Philips was scanned with an 8-ch surface coil and 16-ch volume coil in 3.0 Tesla MR imaging system and the SNR and image uniformity obtained from T1-weighted spin echo(SE) images and T2-weighted fast SE images with the two coils were compared.The images from a volunteer with the two coils were also evaluated.Results The SNR with the 8-ch surface coil is three times higher than the 16-ch volume coil in the region of interest(ROI),which was set in the center of the phantom images.With regard to the spatial variation of sensitivity,the uniformity of 8-ch surface coil is much smaller than that of 16-ch volume coil.Conclusion The images from 8-ch phased array surface coil had higher SNR versus 16-ch volume coil.However,the image uniformity of 8-ch surface coil descends obviously in a larger ROI.