Determination of Blood Glucose by Fluorescence Capillary Analysis Based on Glucose/Glucose Oxidase/H_2O_2/L-tyrosine/ Horseradish Peroxidase Reaction System
LI Qiao-Jing,GAO Xiu-Feng,DU Yun-Dong,CHEN Ting-Mei,LI Yong-Sheng
DOI: https://doi.org/10.3724/SP.J.1096.2011.01181
IF: 1.193
2011-01-01
Chinese Journal of Analytical Chemistry
Abstract:A novel fluorescent reaction system of glucose/glucose oxidase (GOD)/hydrogen peroxide/L-tyrosine/horseradish peroxidase (HRP) combined with a fluorescence capillary analysis (FCA) was established for developing a sensitive glucose enzyme FCA (GE-FCA). After studying the new reaction, the optimum conditions were as follows: concentrations of GOD and HRP were both 2000 U/L, concentration of L-tyrosine was 1.0 mmol/L, pH of phosphate buffer solution was 8.5, reaction temperature was 35 degrees C and reaction time was 20 min. The GE-FCA's linear range was 5.0-140 mu mol/L, relative standard deviation was less than 4.0% and detection limit was 2.65 mu mol/L. The method has been successfully applied for the determination of blood glucose and the recovery was in the range of 98%-104%. Compared with routine methods, merits of the GE-FCA was simple in operation, less in sample volume and dosage of chemical as well as enzymic reagents (only 9 mu L.), wide in linear range, strong in anti- interference ability and low cost, it will be easily popularized.