Abstract:A vibrating mesh nebulizer is a medical device that can break liquid medication into an aerosol of tiny
droplets so that patients can inhale them to treat diseases. The volumetric flow rate of this kind of nebulizer
is determined by many factors, including the orifice geometry, the vibration amplitude and the vibration
mode shape. Thus, many experiments were designed to assess the effects of these different variables on
flow rate and to find the optimal set of device parameters that maximises the flow rate. However, it can
be very expensive to perform such physical experimentation with physical prototypes, which needs different
orifice geometries to be fabricated for different test regimes. The orifice sizes are microscopic with diameters
typically between 3 μm to 5 μm, thus, they are difficult to be fabricated with the exact shape and diameter
that are required. Therefore, virtual prototyping by means of numerical simulation models are required as
the orifice geometries and other factors can be easily changed in the simulation models. This study aims to
provide a numerical simulation model that is experimentally validated. By comparing the simulated flow rates
with the experimentally obtained flow rates, and the simulated droplet diameters with the experimentally
obtained droplet diameters, it was found that the errors are all less than 10% which demonstrates this
numerical simulation model is adequately validated by experiment. Thus, instead of carrying out expensive
experiments, this simulation model can be used for predicting the volumetric flow rates for different prescribed
orifice geometries. Furthermore, this paper also simulates the effects of viscosity and surface tension on the
flow rate, where either viscosity or surface tension increases, the flow rate will decrease.
nanoscience & nanotechnology,engineering, electrical & electronic,instruments & instrumentation,physics, applied