Aerosol Delivery to Simulated Spontaneously Breathing Tracheostomized Adult Model With and Without Humidification

Fai A Albuainain,Jie Li,Fai AlBuainain
DOI: https://doi.org/10.4187/respcare.11495
2024-03-14
Respiratory Care
Abstract:BACKGROUND: Optimal aerosol delivery methods for spontaneously breathing patients with a tracheostomy remain unclear. Thus, we aimed to assess the impact of nebulizer placement, flow settings, and interfaces on aerosol delivery by using a vibrating mesh nebulizer and a jet nebulizer in line with unheated humidification.METHODS: An 8.0-mm tracheostomy tube was connected to the lung model that simulates adult breathing parameters via a collecting filter. Albuterol sulfate (2.5 mg/3 mL) was administered via a vibrating mesh nebulizer and a jet nebulizer, which was placed in line with unheated humidification provided by a large-volume nebulizer, with FIO2 set at 0.28, with gas flows of 2 L/min versus 6 L/min. Nebulizers were placed in line distal and proximal to the lung model by using a tracheostomy collar and a T-piece. Conventional nebulization was tested using a vibrating mesh nebulizer and a jet nebulizer directly connected to the tracheostomy tube bypassing the humidification device. The drug was eluted from the collecting filter and assayed with ultraviolet spectrophotometry (276 nm).RESULTS: During in-line nebulizer placement with unheated humidification, the inhaled dose was 2-4 times higher with a gas flow of 2 L/min than 6 L/min, regardless of nebulizer type, placement, or interface (all P < .05). At 6 L/min, the inhaled dose was higher with proximal than distal placement when using both interfaces, but, at 2 L/min, the inhaled dose was lower with proximal placement. With a jet nebulizer, the tracheostomy collar generated a higher inhaled dose at proximal placement compared with the T-piece, whereas the T-piece resulted in a higher inhaled dose than the tracheostomy collar with distal placement, regardless of the flow settings. Compared with conventional nebulization using a vibrating mesh nebulizer, an in-line vibrating mesh nebulizer with a large-volume nebulizer at 2 L/min had a similar inhaled dose, regardless of nebulizer placement and interface. In contrast, the in-line jet nebulizer was influenced by both placement and interface.CONCLUSIONS: Aerosol delivery with an in-line vibrating mesh nebulizer and jet nebulizer with unheated humidification was affected by nebulizer placement, interface, and gas flow settings.
respiratory system,critical care medicine
What problem does this paper attempt to address?