Study on influencing factors of patients accepting Ventavis ultrasonic atomizing inhalation

Li Jun,Xu Zhuoming,Bian Jun
DOI: https://doi.org/10.3969/j.issn.1009-6493.2009.21.004
2009-01-01
Chinese Nursing Research
Abstract:Objective:To probe into the influencing factors of atomizing generator,tidal volume,drug concentration of ventilator under the mode of pressure-regulated volume control on Ventavis consumption rate,so as to find reasonable parameter setting,to reduce the residual of Ventavis after atomizing inhalation,and to enhance both drug utilization and cost profit.Methods:Mimical breathing circuit was connected with ventilator,test lung,Ventavis,atomizing generator. And Ventavis residual after atomizing inhalation under specific parameter of each group had been observed. Then each experiment had been repeated three times under the same parameter to reduce the error. Results:Through the experiment,we concluded that atomizing generator,tidal volume and body weight of infants have some influences on the Ventavis residual after atomizing inhalation. However,the frequency and time of breathing,the concentration of atomization liquid have no significant influence on the Ventavis residual after atomizing inhalation. Conclusion:It found out influencing factors of the drug consumption rate of Ventavis in atomizing inhalation. It provided some references for enhancing clinical maximum utilization of Ventavis aerosol inhalation and reducing cost at present. It offered some evidence for further exploring mathematical model for studying Ventavis consumption under special condition.