The Design of An Integrated Gradient-Field/Frequency Lock-Shim System for NMR Spectrometer

Hui-wang WANG,Zheng-gang LI,Wen-ping MAO,Qing-jia BAO,Zao LIU,Chao-yang LIU
DOI: https://doi.org/10.3969/j.issn.1000-4556.2012.04.002
2012-01-01
Abstract:The design of an integrated gradient-field/frequency lock-shim system for NMR spectrometer is presented.The integrated architecture is realized by combining embedded ethernet,field programmable gate array(FPGA),external memory interface(EMI),controller area network(CAN) bus and high-speed serial bus technologies,making the system compact and reliable.The system consists of three parts:(1) the field/frequency lock subsystem which ensures the stability of the magnetic field,(2) the shim subsystem which guarantees the consistency of the magnetic field,and(3) the gradient amplifier which generates the gradient field pulse.Due to its flexibility,the design can easily be adapted for different NMR spectrometers.The feasibility of the integrated system is illustrated by tests of the shim,lock and gradient outputs on a 11.7 T NMR spectrometer.
What problem does this paper attempt to address?