The Equivalent Circuit Model and Changes of Impedance Parameters during Impedance Microbial Detection

Zhang Aiping,Wu Dan,Hou Shijuan,Tang Jiani,Liu Donghong
DOI: https://doi.org/10.3969/j.issn.1009-7848.2011.02.029
2011-01-01
Abstract:In this paper,the equivalent circuit of the test tube during impedance microbiological detection was analyzed by the equivalent circuit method,obtaining that the total impedance containing the constant conductance,the conductance changed exponentially with frequency and the imaginary part of impedance changed exponentially with frequency through analyzing the Bode diagram.Combined with the Nyquist diagram the equivalent circuit of the test tube was the constant phase angle element(CPE) in series with the resistor R.Finally,getting the equivalent circuit model using matlab:Z = 11.423+1.3704×104×ω-0.7767×cos(0.7767π2)-j1.3704×104×ω-0.7767×sin(0.7767π2).Meanwhile,this paper also studied the changes and the rate of changes of the electric double-layer capacitance(Cs),the conductance(G),the total impedance(Z),the impedance angle(θ) during microbial growth-Cs,G increases,Z,θ decreases;the trend of the changes rate of Cs,G,Z,θ was the same,but the amplitude of the changes rate were different-Cs was the maximum,while Z was minimum.