Impedimetric detection of yeast and mold

B.J. Schaertel,Nora Tsang,Ruth Firstenberg-Eden
DOI: https://doi.org/10.1016/0740-0020(87)90031-1
IF: 6.374
1987-04-01
Food Microbiology
Abstract:An automated impedance test was developed for the rapid detection of yeast and mold. Various mycological media were evaluated using the capacitance (C) and conductance (G) components of the impedance signal. C and G were compared as to their effectiveness in monitoring microbial growth-induced changes in the media. Similarly, agar vs. broth media were compared. 0·1 ml aliquots of yeast and mold food isolates were loaded into duplicate Bactomatic module wells pre-filled with media (0·5 ml agar; 1·0 ml broth) to give 102–103 cells well−1. Each organism-medium combination was monitored by C and G at 28°C. C measurement was more effective than G measurement was. C monitoring was characterized by a sharp acceleration (i.e., detection) in the resulting curves as the cell concentration reached 104–105 cells ml−1. Conversely, weak, irregular responses to G obscured detections. The agars, particularly Bactomatic Yeast Medium and Wort Agar, gave the best results. Broth curves exhibited ambiguous detections and less percent change/h. An impedimetric test detects mold in 1–2 days compared to a 5 day standard plate count. Selection of appropriate media and signal (C) optimizes impedimetric results for yeast and mold. An agar surface supports mold and yeast growth better than does a broth.
food science & technology,biotechnology & applied microbiology,microbiology
What problem does this paper attempt to address?