Application of CT Perfusion Imaging on Transjugular Vein Intrahepatic Portal-Systemic Shunt

李家平,陈伟,黄勇慧,王海林,杨建勇
DOI: https://doi.org/10.3969/j.issn.1001-9324.2008.05.027
2008-01-01
Abstract:Objective To evaluate the hepatic perfusion changes before and after transjugular vein intrahepatic portal-systemic shunt(TIPSS)by means of spiral CT perfusion imaging.Materials and Methods CT perfusion imaging was performed in fifteen patients with cirrhosis and portal hypertension 2 days before TIPSS and one week after TIPSS.The hepatic perfusion parameters before and after TIPSS were compared.Results Prior to TIPSS,portal venous perfusion(PVP),hepatic arterial perfusion(HAP),total liver perfusion(TLP),and hepatic arterial perfusion index(HPI) was(0.58±0.23) ml·min-1·ml-1,(0.14±0.13) ml·min-1·ml-1、(0.72±0.17) ml·min-1·ml-1,and(24.0±10.2)%,respectively.After TIPSS,PVP,HAP,TLP,and HPI was(0.15±0.04)ml·min-1·ml-1,(0.28±0.05) ml·min-1·ml-1,(0.43±0.07) ml·min-1·ml-1 and(64.1±13.9)%,respectively.There was significant difference in the hepatic function,blood ammonia and portal free pressure between pre-TIPSS and post-TIPSS.Conclusion Spiral CT perfusion imaging could reveal the hemodynamic changes of the liver before and after TIPSS.