Solid phase microextraction and GC for simultaneous measurement of aniline and pyridine in water

Zhong Ming,Sun Cheng,Li Xiang
DOI: https://doi.org/10.3969/j.issn.1001-3865.2006.03.021
2006-01-01
Abstract:Solid-phase microextraction(SPME) using a PDMS/DVB fiber and the subsequent gas chromatography using a flame ionization detector(GC-FID) were employed for simultaneous measurement of aniline and pyridine in water samples.The extraction performed at 15 ℃ for 30 minutes at a pH of 7.0 would result in the largest GC signal.The detection limit ranged from 0.050 to 0.065 mg/L,and the relative standard deviation was 6.44% ~8.67%.The SPME-GC method is simple,fast,precise,sample efficient and solvent-free.
What problem does this paper attempt to address?