Pressure Measurement System Using Surface Wave Based on TDC

Ling Zhangwei,Zhang Hongjian,He Qing
DOI: https://doi.org/10.13382/j.jemi.2008.03.005
2008-01-01
JOURNAL OF ELECTRONIC MEASUREMENT AND INSTRUMENT
Abstract:Nondestructive pressure measurement based on surface wave usually requests accurate measurement of propagation time of surface wave in the medium.A dual sing-around measurement system was developed aiming at the reference type and calculation modifying pressure type measurement methods.Using this system,not only the propagation time of surface wave can be measured using sing-around method,but also the propagation time difference between two surface waves can also be detected.Experiment results show that the root mean square error is 0.0077ns for the propagation time measurement;the maximal absolute error is not greater than 0.07MPa for reference pressure measurement and the maximal absolute error is not grater than 0.12MPa for calculation modifying pressure measurement.
What problem does this paper attempt to address?