Determination of Daunomycin at a Novel COOH/indium Tin Oxide Ion Implantation-Modified Electrode

DongMei Gao,JingBo Hu,Meng Yang,QiLong Li
DOI: https://doi.org/10.1016/j.ab.2006.06.008
IF: 2.9
2006-01-01
Analytical Biochemistry
Abstract:A novel COOH+ ion implantation-modified indium tin oxide (COOH/ITO) electrode was prepared for the first time and applied for determination of daunomycin (or daunorubicin [DNR]). The electrode showed higher catalytic activity than bare ITO electrode with good reproducibility and stability. The determination condition of linear voltammetry was optimized. A calibration curve was obtained over the range 2.0 x 10(-7) to 5.0 x 10(-6)mol/L, with a correlation coefficient of 0.9909 and a limit of detection of 1.0 x 10(-7)mol/L. The selectivity of the electrode was illustrated by determination of DNR in samples prepared of urine. X-ray photoelectron spectroscopy (XPS) analysis results showed that the implanted COOH+ maintained characteristics of organic group -COOH. A field emission-scanning electron microscope (FSEM) result showed that the implanted surface caused defects and partial dislocations and formed many active centers.