Content Determination of Posaconazole Eye Drops by HPLC

WU Wei,HE Mei-feng,GUO Ze-li,ZHONG Jian-wen
DOI: https://doi.org/10.6039/j.issn.1001-0408.2013.25.23
2013-01-01
Abstract:OBJECTIVE: To establish a method for the content determination of Posaconazole eye drops. METHODS: HPLC method was adopted. The determination was performed on Diamonsil C18 column with mobile phase consisted of acetonitrile-water (60 ∶ 40) at the flow rate of 1.0 ml/min. The detection wavelength was set at 262 nm and column temperature was 25 ℃. RESULTS: The calibration curve of posaconazole was linear within the concentration range of 8.181-81.81 μg/ml (r=0.999 9) with an average recovery of 99.4% (RSD=0.77%, n=9). CONCLUSIONS: The method is simple, rapid and accurate, and it can be used for the quality contro1 of Posaconazole eye drops.
What problem does this paper attempt to address?