Effect of Field Strength on MR Imaging of Knee

LIU Wei,JIANG Quan-sheng,DU Xiang-ke
DOI: https://doi.org/10.3969/j.issn.1674-1633.2009.08.060
2009-01-01
Abstract:Objective To assess the effect of field strength on magnetic resonance imaging of knee.Methods 16 healthy volunteers were imaged at two similarly equipped MR imagers of the same generation and from the same manufacturer.The same imaging sequences were used with identical parameters and without repetition time correction for field strength.Imaging was performed in the same anatomic locations: the knee.Quantitative image analysis involved calculation of signal-to-noise ratio,contrast-to-noise ratio and relative contrast.Results SNR(signal-to-noise ratio) and CNR(contrast-to-noise ratio) increased nonlinearly with field strength,3.0T was judged superior to 1.5T images(P<0.05),relative contrast was not as dependent on field strength.Image quality was judged to be equivalent at 1.5T and 3.0 T.Conclusion Excellent image quality was obtained with two kinds of field strengths.Compared to 1.5T,SNR and CNR were promoted at 3.0T.However,relative contrast had no significant differences.