Comparison of cartilage signal intensity in different magnetic resonance imaging

LU Yong,DING Xiao-yi,SONG Wei-feng,DU Lian-jun,YAN Ling,JIANG Hao,LIU Jian-jun,CHEN Ke-min
DOI: https://doi.org/10.3969/j.issn.1671-2870.2007.03.014
2007-01-01
Abstract:Objective To investigate the ability and limitation of clinical cartilage magnetic resonance imaging (MRI) through signal-noise ratio (SNR) and contrast-noise ratio (CNR) analysis. Methods SE T1W,FSE T2W,STIR,FS-PD,3D-FS-SPGR and T2*W GRE imaging were performed in 12 adult cadaveric knees with GE Signal 1.5T unit. The signal intensity (SI) of the cartilage in femoral condyle,femoral metaphysic ossium,gastrocnemius and background noise was measured in the imaging located in femoral condyle fossa. Then the cartilage SNR and CNR between cartilage and medulla ossium,synovia and myoideum were calculated. Results The cartilage SNR in 3D-FS-SPGR and FS-PD imaging was higher than the values in the other sequences (P<0.001); the cartilage / medulla ossium CNR in 3D-FS-SPGR imaging was the highest in all the MR sequences(P=0.0123); the cartilage/synovia CNR and the cartilage/ myoideum CNR in 3D-FS-SPGR and FS-PD imaging were higher than the values in the other sequences (P<0.001). Conclusions 3D-FS-SPGR and FS-PD imaging are the optimal sequences for articular cartilage MRI.