Micro-electrospray with stainless steel emitters.

Wenqing Shui,Yanling Yu,Xuejiao Xu,Zhenyu Huang,Guobing Xu,Pengyuan Yang
DOI: https://doi.org/10.1002/rcm.1083
IF: 2.586
2003-01-01
Rapid Communications in Mass Spectrometry
Abstract:The physical processes underlying micro-electrospray (micro-ES) performance were investigated using a stainless steel (SS) emitter with a blunt tip. Sheathless micro-ES could be generated at a blunt SS tip without any tapering or sanding if ESI conditions were optimized. The Taylor cone was found to shrink around the inner diameter of the SS tubing, which permitted a low flow rate of 150 nL/min for sheathless microspray on the blunt tip (100 mum i.d. x 400 mum o.d.). It is believed that the wettability and/or hydrophobicity of SS tips are responsible for their micro-ES performance. The outlet orifice was further nipped to reduce the size of the spray cone and limit the flow rate to 50-150 nL/min, resulting in peptide detection down to attomole quantities consumed per spectrum. The SS emitter was also integrated into a polymethylmethacrylate microchip and demonstrated satisfactory performance in the analysis and identification of a myoglobin digest. Copyright (C) 2003 John Wiley Sons, Ltd.