Cervical Zone of Cortical Bone with the Obvious Resorption of the Bone

SONG Tao,CAI Guoping,XIU Peicheng
DOI: https://doi.org/10.3969/j.issn.1008-6358.2011.02.030
2011-01-01
Abstract:Objective:To observe the change of he stress on implant and bone interface and the critical point of the maximum security angle under the condition of different angles of the implants direction and dental crowns long axis of the maxillary central incisor so as to guide the implant designs.Methods: The finite element method was used to analyze the change of the stress on implant and bone interface under different angles of the implants direction.We found the ideal position of the long axis of the maxillary central incisors with combining clinical cases and cephalographs.The stress change of implant and bone interface and the marginal bone loss around implants are both analyzed when the angle between the implant and the ideal location of the incisors changes.Results: When the abutment angulations of the maxillary central incisor were between 0 to 20,the marginal bone resorption and the implant-bone-interface stress were normal.But when the abutment angulations of the maxillary central incisor increased to 20 or 25,the implant stress increases sharply.Peak tensile stresses were focused in labial.Conclusions: It is suggested that the alveolar bone remodeling be done and the change of the abutment angulation.
What problem does this paper attempt to address?