Quality Control Analysis and Discussion on Noninvasive Blood Pressure Detection of ECG Monitor
SHI Min-chao,JIN Wei,QIAN Zheng-ying
DOI: https://doi.org/10.3969/j.issn.1003-8868.2012.06.051
2012-01-01
Abstract:Objective To control the performance situation of ECG monitors in use through the quality control work.Methods Fluke's professional quality control equipment was used to detect the ECG monitors' noninvasive blood pressure and the results were analyzed deeply.Results The total pass rate of the ECG monitors' noninvasive blood pressure measurement was high.There were differences among the maximum error ranges of ECG monitors from different brands or models,and also among the different measurement ranges.Conclusion The test results are analyzed from 3 aspects including brand or model,values differences among different measurement ranges and service age differences,which has great significance to identify the factors that affect the noninvasive blood pressure test results,and to improve the pass rate of the instrument.[Chinese Medical Equipment Journal,2012,33(6):114-115]