Accuracy Study of Digital Blood Pressure Monitor in Clinic Measurement

Huaqing Zhang,Chao Huang,Yiwen Wang,Shuming Ye
DOI: https://doi.org/10.3969/j.issn.1002-2376.2016.13.023
2016-01-01
Abstract:To achieve the standardization, informatization and automation management of the hospital, the certified digital blood pressure (BP) monitors were used to measure BP instead of traditional auscultatory method. However, since the BP measurement was easily effected by several uncertain factors, the results may be unreliable and due to the measurement could not be retrospected effectively, some hypertension patients may be missed or misdiagnosed. To study the accuracy of the clinic measurement by digital BP monitor, this paper randomly selected 640 patients from the hospital to compare with the other ambulatory BP monitor, which could be re-diagnosed after BP measurement. Using the evaluation method proposed by the British Hypertension Society (BHS), the percentage of the systolic and diastolic BP in the ranges of ≤5 mmHg, ≤10 mmHg, ≤15 mmHg, was 41%, 69%, 86% and 49%, 81%, 93%,respectively. The result shows that the accuracy could not satisfy the recommendation of the BHS standard in clinic application. Therefore, to achieve the accurate and effective management of BP database, we suggest that besides the standardization of the BP measurement procedure, it is recommended to establish a system to retrospect the BP results effectively.