Comparison of portable real-time remote electrocardiogram monitor with ambulated electrocardiogram monitoring

Ze Feng,Sun Yujie,Li Ding,Zhang Haicheng
DOI: https://doi.org/10.3969/j.issn.1005-0272.2013.02.007
2013-01-01
Abstract:Objective To investigate the effectiveness, reliability and accuracy of waves recorded of the portable real-time remote electrocardiogram monitor, compared with ambulatory electrocardiogram monitoring. Methods Thirty-six hospitalized patients with arrhythmia were equipped with both the portable real-time ECG monitor (AKM-N1, Auko) and three-leading Holter (MIC-Dream V5.3.60.10s.3).for 1 hour, and then compare the ECG data of each patients. Results There is no statistically significant difference (p>0.05) in P wave duration, PR interval, QRS complex duration, QT interval and RR interval between the two groups. Interference and pseudo rate of the ECG detector group is lower then that in the Holter group. There is no significant difference (p<0.05) in detectable rate of arrhythmia between the two groups. Conclusions Compared with the ambulatory electrocardiogram monitoring, the portable real-time remote electrocardiogram monitor could be used in detecting cardiac arrhythmias.
What problem does this paper attempt to address?