Comparative Study of HPLC and UV Method for the Determination of Sorafenib

Zhang Hong,Zhang Fuming,Yan Shijun
DOI: https://doi.org/10.3969/j.issn.1006-4931.2012.16.022
2012-01-01
Abstract:Objective To quantify the sorafenib using both HPLC and UV method.Methods HPLC analyses were carried out on an Inertsil ODS-3 column(250 mm×4.6 mm,5 μm)with a mobile phase of ammonium acetate buffer(50 mmol/L)-acetonitrile(28 ∶72)at the flow rate of 1.0 mL/min and a 265 nm detection.The column temperature was 40 ℃and injected sample amount was 50 μL.The UV method with methanol as solvent to prepare a sample has been developed using UV detection at 265 nm.Results HPLC method was validated and found to be linear in the range of 0.5-40 μg/mL(r=0.999 99);The average recovery was 102.92%with RSD of 1.30%(n=9).Meanwhile,the linear range of sorafenib was 0.1-20 μg/mL(r=0.999 9)by UV method.The result of UV method was 1.067 times higher than HPLC method.Conclusion The UV method is not appropriate for the determination of sorafenib.The HPLC method with good specificity,stability,precision and recovery rate is appropriate for the determination of sorafenib.
What problem does this paper attempt to address?