Determination of the Plasma Sirolimus Concentrations in Patients after Renal Transplantation by HPLC Method

Hong Xu
2004-01-01
Abstract:Objective: To monitor plasma concentrations of sirolimus in patients after renal transplantation. Methods: HPLC method was used with a column of YMC-Pack ODS-A (150 mm×4.6 mm, 5 μm) and a guard column of Alltima C 18 (7.5 mm×4.6 mm,5 μm). The mobile phase was a mixture of acetonitrile-tetrahydrofuran-H 2O (55∶5∶40). Sirolimus was detected at 278 nm and 32-desmethoxysirolimus was used as the internal standard. Results: The retention time of sirolimus and 32-desmethoxysirolimus were about 10.1 min and 12.1 min, respectively. A good linearity was shown over the concentration range of 2.49-76.36 ng /ml. The method recoverly rates were between 99.93% and 105.9%. The intra-day or inter-day RSD were less than 7.40%. Conclusion: This HPLC method is sensitive, simple and accurate.