Measurement of ciclosporin in whole blood by HPLC

Wen Xiushu,Shan Lvhu,Tao Zhihua,Yang Jianrong,Zhu Yanying
DOI: https://doi.org/10.3969/j.issn.1007-4406.2006.06.011
2006-01-01
Abstract:AIM To establish a HPLC method for determination of ciclosporin.METHODS Blood samples(1 mL)were deproteinzed by addition of 1 mL of a mixture of 800 mL of methanol and 200 mL of aqueous zinc sulfate solution(0.4 mmol·L~(-1)).Then the supernatant liquid was pipetted into C_(18)(10 mm×3.9 mm) column.First,samples were injected into the HPLC system and transferred through the extraction column by water 2.0 mL·min~(-1) for 2 min;second,the ciclosporin was eluted by methanol.The separation was achieved using an C_(18)(250mm×4.2mm) column with a mobile phase of acetonitrilemethanol-water(60∶10∶30,V/V/V),then UV detection was performed at 210 nm.RESULTS In whole blood the liner range was 62.5-2 000 μg·L~(-1),the minimal detectable concentration of this method was 40(μg·)L~(-1),the precision of method was 90%-95%,the extraction recovery of method was more than 80%(n=5),the RSD was within 3.25%-4.12% within day,and within 3.37%-6.87% between days.CONCLUSION The method is simple and could be applied to examine the clinical samples.