A linear time-of-flight mass analyzer for thermal ionization cavity mass spectrometry

David M. Wayne,Wei Hang,Diane K. McDaniel,Robert E. Fields,Eddie Rios,Vahid Majidi
DOI: https://doi.org/10.1016/S0584-8547(01)00207-5
2001-01-01
Abstract:In this paper we report the basic design characteristics, typical operating parameters, and isotope ratio performance of an orthogonal acceleration linear thermal ionization cavity time-of-flight mass spectrometer (TIC-TOFMS). The present system is capable of mass resolution of 750–850 (FWHM) over a wide range of masses, and can generate and analyze multi-element spectra from sub-μg samples (solids and solution residues) in <30–45 min. The optimum precision (1σ) of isotope ratios determined from 60–80 spectra (each the average of 600 individual spectra) is 0.2–0.4% R.S.D., and is limited by the instrument drift, dead time and the data acquisition and processing capabilities of the 8-bit digital oscilloscope used to collect the data. Isotope ratio accuracy (1σ, per mass unit) for major isotopes is typically <±1.0%.