Influence of systematic discrepancy in coagulation tests on clinical decision

WANG Hai,WANG Cheng-bin,QI Xiao-wei,LI Jian
DOI: https://doi.org/10.3877/cma.j.issn.1674-0785.2012.15.101
2012-01-01
Abstract:Objective To investigate discrepancies of coagulation parameters in different systems,and try to standardize or calibrate these index.Methods 247 samples were collected from patients,to observe difference of PT/INR,APTT,and FIB tested in STA-R,CA7000 and ACL-Advance,and then analyzed by Student-Newman-Keuls test.INR data was divided into 3 groups:<2,2-3,and>3 to make further analysis by nonparameter test.Results There were significant difference in PT,APTT and FIB(F=32.570,P<0.01;F=134.076,P<0.01;F=99.873,P<0.01)respectively.Both PT and APTT from STA-R and CA-7000 system were correlated(r=0.984 and 0.944,all P<0.01).Nonparameter test showed no difference(χ2=7.583,P<0.05)of INR in three levels,when system-specific ISI was used.Fibrinogen measured by Clauss were different from that of PT-derived(F=99.873,P<0.01).Moreover ACL Advance and STA-R showed good correlation(r=0.909,P<0.01)in Fib,especially in low fibrinogen level.Conclusions It is difficult to minimize the difference merely by a correct formula,although good correlation lies between systems.We should continue to emphasize on the reagent quality and traceability,and pay more attention on system specialty to keep accuracy of coagulation tests.