Enhancing the Standardization of International Normalized Ratio with Lyophilized Plasma Calibrants

李健,丛玉隆,王成彬,谭笑红,陈伟,邸平
DOI: https://doi.org/10.3969/j.issn.1009-6213.2008.06.008
2008-01-01
Abstract:Objective To reduce the variability of international normalized ratio(INR) reporting prothrombin time(PT) amoung laboratories or detecting systems,evaluate the applicability of lyophilized PT-multi calibrator in multi-system calibrations.Methods Eastablishing a local ISI and PT-INR calibration curve for reagent-and equipment-special system with PT-multi calibrator,respectively.Samples were collected and divided into 3 groups corresponding to their INRs,determined by diffirent conbinations of certain equipment and reagent recommanded by manufacture or subsititude reagent.Results The linear correlation between any two systems with similar method was good enough,but the mean values still have statistical difference.With the help of PT-INR curve,gaps between systems were minimized(P0.05),especially when substitute reagent is used.Conclusion Calibration with special calibrator is necessary to improve the INR accuracy and inter-laboratory precision,when reagents not assigned by the manufacture are used or analyzing system has any change.