Determination of Sodium Ferulate by a Flow-Injection Chemiluminescence Inhibition Method

蔡卓,江彩英,杜良伟,赵静,莫丽君
DOI: https://doi.org/10.3969/j.issn.1000-0720.2010.04.020
2010-01-01
Abstract:A novel flow injection chemiluminescence (FI-CL) method for the determination of sodium ferulate was developed.It was observed that a strong chemiluminescence (CL) occurred when methylene blue solution was mixed with fenton reagent in an acidic medium,and the CL intensity greatly decreased with the addition of sodium ferulate into the reaction system.The decreased CL intensity was linear with the concentration of sodium ferulate in the range of 4.5×10-6~4.5×10-5 mol/L under the optimum experimental conditions.The detection limit was 7.0×10-7 mol/L,and the relative standard deviation (RSD) was 0.8% for 4.8×10-6 mol/L sodium ferulate (n=11).This proposed method had been successfully applied to the determination of sodium ferulate in tablets.
What problem does this paper attempt to address?