New detection schemes in liquid chromatography

Edward S. Yeung
DOI: https://doi.org/10.1016/0731-7085(84)80076-X
IF: 3.571
1984-01-01
Journal of Pharmaceutical and Biomedical Analysis
Abstract:A micropolarimeter interfaced to a liquid chromatograph is shown to be suitable for selective monitoring of the optically-active components in complex samples. When an optically-active eluent is used, indirect determination of even optically-inactive materials is possible, down to the level of 10 ng of an injected component. If a second chromatogram is obtained using the racemic analogue of the optically-active eluent, quantitation can be achieved without standards and without prior analyte identification. This concept is also applicable to the refractive index detector, the absorption detector and the conductivity detector in the special case of ion chromatography, and the ultrasonic detector in gas chromatography.