Design of Micro Current Detection Circuit for FAIMS

YAN Shu-jiang,TANG Fei,WANG Xiao-hao,WANG Fan,YANG Tao
DOI: https://doi.org/10.3724/sp.j.1187.2011.00711
2011-01-01
Abstract:In order to solve the problems of the capacitance integral circuit which cannot meet the requirement of fast detection for FAIMS,a new design of circuit for micro current detection is introduces.Proper components and circuit design are adopted to amplify pA current up to mV order of magnitude with various noise sources of pre-circuit taken into consideration.In the experiment,the current signal is collected and processed by NI acquisition board under the control of LabVIEW,and the result shows that the measurement accuracy of the circuit is lower than 0.1 pA and the sampling frequency can reach 15 Hz.In the real FAIMS test,the system background noise acquired by the circuit is 0.2 pA and the FAIMS spectra of ethanol is obtained under the condition of carrier gas flow of 0.8 L/min and the compensation voltage from-4 V to 6 V at 0.1 V sweeping step.The circuit can meet the requirements of FAIMS well.
What problem does this paper attempt to address?