Investigation on extreme ultraviolet photon-counting imaging detector with position-sensitive anode

NI Qi-liang,LIU Shi-jie,CHEN Bo
DOI: https://doi.org/10.3969/j.issn.2095-1531.2009.01.006
2009-01-01
Chinese Journal of Optics and Applied Optics
Abstract:A prototype of two-dimensional photon-counting imaging detector with position-sensitive anode in Extreme Ultraviolet(EUV) region was developed for applying to a moon-based EUV imager.The detector mainly is composed of Microchannel Plate(MCP) stacks operating in pulse-counting mode,wedge-strip position-sensitive anode and correlative analog and data processing circuit.A three electrode wedge-strip position-sensitive anode with 1 mm period and 30 mm-diameter was designed and fabricated,and a front-end analog and digital circuit with maximum rate of 200 kHz also was developed.The operating characteristics of the detector prototype including background rate,pulse height distribution and spatial resolution were measured,the measured results show its spatial resolution is 0.28 mm,which can satisfy the requirement of moon-based EUV imager on the spatial resolution.
What problem does this paper attempt to address?