Application of intraoperative laparoscopic ultrasonography in the cool-tip radiofrequency ablation of hepatic carcinoma

QIU Jun-lan,ZHOU Tie,KONG Wen-tao,JIANG Zhi-ming,SHEN Yi,ZHANG Wei-wei
DOI: https://doi.org/10.3877/cma.j.issn.1672-6448.2012.09.019
2012-01-01
Abstract:Objective To evaluate the feasibility and efficacy of intraoperative laparoscopic ultrasonography(ILU) in the radiofrequency ablation(RFA) of hepatic carcinoma.Methods Thirty-one patients with hepatic carcinoma were enrolled in this study.In the process of RFA,ILU was applied to guide the RFA needle and to monitor the process of RFA in real-time.Results Under the guidance of ILU,the percutaneous RFA in all patients proceed smoothly.Seven additional masses at the liver diaphragmatic that were missed in preoperative ultrasound but reported by preoperative CT or MRI top were detected.Three small lesions were not detected in preoperative imaging.A false positive focal lesion was ruled out accurately.Thirty-one cases of liver cancer patients underwent laparoscopic surgery without intraoperative death,postoperative infection,haemorrhage or other complications.Five cases had transient fever,four had nausea and three had mild abdominal pain.Patients were followed up for six months with no recurrence.Conclusions ILU has certain advantages to routine ultrasound in observing conventional blind area and detecting missed lesions preoperatively.ILU is also helpful for assisting the percutaneous RFA treatment in patients that are unable to detect or locate by preoperative ultrasound and intolerant to routine surgery.
What problem does this paper attempt to address?