HPLC determination of talinolol in human plasma

YAN Miao,ZHANG Jin-jiao,LI Huan-de,LI Lan-fang,ZHENG Mei
DOI: https://doi.org/10.13286/j.cnki.chinhosppharmacyj.2013.16.022
2013-01-01
Abstract:OBJECTIVE To establish a rapid and sensitive HPLC-UV method for determination of talinolol in human plasma and assessment of the pharmacokinetics of talinolol tablets.METHODS The analytes in plasma were extracted by tert-butyl methyl ether as the solvent after samples had been spiked with propranolol.The analytical column was packed with a Phenomenex C18(4.6 mm×250 mm,5 μm).The mobile phase was acetonitrile-methyl alcohol-10 mmol·L-1ammonium acetate acid(25∶10∶65),with a flow rate of 1.0 mL·min-1.The UV detection wavelength was 248 nm and the temperature was 38 ℃.RESULTS Excellent linear relationship was obtained from the range of 8 ng·mL-1 to 500 ng·mL-1(r2=0.999 9).The limit of detection(LOD) was 8.0 ng·mL-1 for talinolol.The average extraction recovery was above 82.3%,and the methodology recovery was above 106.1%.Talinolol was fitted demands of biological sample.CONCLUSION The method is accurate,sensitive and simple for the pharmacokinetiec study of talinolol.
What problem does this paper attempt to address?