Volumetric Measurement of Pulmonary Nodules at Low-Dose Chest CT: Effect of Reconstruction Setting on Measurement Variability

Ying Wang,Geertruida H. de Bock,Rob J. van Klaveren,Peter van Ooyen,Wim Tukker,Yingru Zhao,Monique D. Dorrius,Rozemarijn Vliegenthart Proença,Wendy J. Post,Matthijs Oudkerk
DOI: https://doi.org/10.1007/s00330-009-1634-9
IF: 7.034
2009-01-01
European Radiology
Abstract:Objective To assess volumetric measurement variability in pulmonary nodules detected at low-dose chest CT with three reconstruction settings. Methods The volume of 200 solid pulmonary nodules was measured three times using commercially available semi-automated software of low-dose chest CT data-sets reconstructed with 1 mm section thickness and a soft kernel (A), 2 mm and a soft kernel (B), and 2 mm and a sharp kernel (C), respectively. Repeatability coefficients of the three measurements within each setting were calculated by the Bland and Altman method. A three-level model was applied to test the impact of reconstruction setting on the measured volume. Results The repeatability coefficients were 8.9, 22.5 and 37.5% for settings A, B and C. Three-level analysis showed that settings A and C yielded a 1.29 times higher estimate of nodule volume compared with setting B ( P = 0.03). The significant interaction among setting, nodule location and morphology demonstrated that the effect of the reconstruction setting was different for different types of nodules. Low-dose CT reconstructed with 1 mm section thickness and a soft kernel provided the most repeatable volume measurement. Conclusion A wide, nodule-type-dependent range of agreement between volume measurements with different reconstruction settings suggests strict consistency is required for serial CT studies.
What problem does this paper attempt to address?