High-Performance Liquid Chromatography-Electrospray Ionization-Mass Spectrometric Determination of Emedastine Difumarate in Human Plasma and Its Pharmacokinetics

Yuan Tian,Zunjian Zhang,Jiabi Liang,Wei Li,Yun Chen
DOI: https://doi.org/10.1016/j.jpba.2008.03.015
IF: 3.571
2008-01-01
Journal of Pharmaceutical and Biomedical Analysis
Abstract:A simple, rapid, sensitive and selective liquid chromatographic–tandem mass spectrometric (LC–MS/MS) method was developed and validated for the quantification of rupatadine in human plasma using estazolam as internal standard (IS). Following liquid–liquid extraction, the analytes were separated using a mobile phase of methanol–ammonium acetate (pH 2.2; 5mM) (50:50, v/v) on a reverse phase C18 column and analyzed by a triple-quadrupole mass spectrometer in the positive ion and multiple reaction monitoring (MRM) mode, m/z 416→309 for rupatadine and m/z 295→267 for the IS. The assay exhibited a linear dynamic range of 0.1–100ng/ml for rupatadine in human plasma. The lower limit of quantification (LLOQ) was 0.1ng/ml with a relative standard deviation of less than 20%. Acceptable precision and accuracy were obtained for concentrations over the standard curve range. The validated LC–MS/MS method has been successfully applied to study the pharmacokinetics of rupatadine in healthy volunteers.
What problem does this paper attempt to address?