Voltammetric Determination of Cadmium(ii) Using a Chemically Modified Electrode

Shengshui Hu,Kangbing Wu,Hongchao Yi,Xuan Dai,Dafu Cui
DOI: https://doi.org/10.1007/s002160100700
2001-01-01
Abstract:An 1-(pyridylazo)-2-naphthol modified glassy carbon electrode has been investigated as sensor for the measurement of trace levels of Cd2+. Cd2+ is deposited on the surface of a PAN modified glassy carbon electrode at -1.10 V (vs. SCE) via forming Cd2+-PAN and subsequent reduction at the electrode. In the following step, Cd-PAN is oxidized, and voltammograms are recorded by scanning the potential in a positive direction. Calibration plots were found to be linear in the range 2 x 10(-8) mol/L to 8 X 10(-7) mol/L. The detection limit was 5 x 10(-10) mol/L, and the coefficient of variation, determined on one single electrode at a concentration of 5 x 10(-7) mol/L, was calculated to be 3.2% (n = 5). Using this new kind of modified electrode. trace levels of Cd(II) in water samples were determined; the average recovery was calculated to be 98.78%.
What problem does this paper attempt to address?