A Blood Glucose Detecting System Based on Three Wavelengths

XIAO Jun,WANG Long,LUO Qing-ming,GONG Hui
DOI: https://doi.org/10.3321/j.issn:1005-0086.2007.09.033
2007-01-01
Abstract:A system based on three-wavelength in near-infrared range was designed to detect the blood glucose concentration.The system is composed of a light source,a chopper,a detector and a lock-in amplifier.1 200 nm,1 350 nm,1 610 nm were chosen in this system.The detector contained ring source and round photodiode.And the distance of source-detector was 3~5 mm which is derived by theoretic analysis.The result indicated that the ratio of absorption difference D depended on the glucose level linearly,and the relative coefficient was 0.972.When using partial least square(PLS) to predict,the root mean square error of prediction (RMSEP) was 17.08 mg/dL and the relative coefficient was 0.997.