A Non-invasive Blood Glucose Detection System Based on Photoplethysmogram with Multiple Near-infrared Sensors

Kai Du,Jing Li,Zhiyi Huang,Houbing Song,Yuanting Zhang,Yuan Zhang,Zhen Mei
DOI: https://doi.org/10.1109/JBHI.2024.3443317
2024-08-14
Abstract:Recent advancements in non-invasive blood glucose detection have seen progress in both photoplethysmogram and multiple near-infrared methods. While the former shows better predictability of baseline glucose levels, it lacks sensitivity to daily fluctuations. Near-infrared methods respond well to short-term changes but face challenges due to individual and environmental factors. To address this, we developed a novel fingertip blood glucose detection system combining both methods. Using multiple light sensors and a lightweight deep learning model, our system achieved promising results in oral glucose tolerance tests. A total of 10 participants were involved in the study, each providing approximately 700 data segments of about 10 seconds each. With a root mean squared error of 0.242 mmol/L and 100% accuracy in zone A of the Parkes error grid, our approach demonstrates the potential of multiple near-infrared sensors for non-invasive glucose detection.
What problem does this paper attempt to address?