Deadtime Correction of Several In-Situ HPGe Γ Spectrometers

陈伟,冯天成,晏林,苏川英,吴睿,龙斌,冯元举
IF: 1.279
2010-01-01
Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms
Abstract:Deadtime corrections of three in-situ HPGe γ spectrometers were studied by experimental method of interference-method with strong γ source.The linear relationships between the percent deadtime and the correction coefficients were fitted by the least-square-method.The comparisons between the experimental results and the automatic correction values of full-energy-peak count rates were performed.The results show that the maximum deviation of the automatic correction values relatively to the experimental results is less than 4.4%.It indicates that in the deadtime range of 1.13%-52.95%,the automatic correction results of deadtime by these spectrometers themselves are exact and the deadtime correction techniques of modern γ spectrometers are efficacious,which can be used in the data analysis of γ spectra.
What problem does this paper attempt to address?