Study on After-Pulses of FPMTs
Lingyue Chen,Lishuang Ma,Guorui Huang,Zhehao Hua,Muchun Jin,Zhen Jin,Sen Qian,Ling Ren,Shuguang Si,Jianning Sun,Qi Wu,Xingchao Wang,Yifang Wang,Zhi Wang,Ning Wang,Kai Wu,Haoda Zhang
DOI: https://doi.org/10.1088/1748-0221/19/10/t10004
2024-01-01
Journal of Instrumentation
Abstract:Abstract The small-sized Micro-Channel Plate Photomultiplier Tube (MCP-PMT), also called Fast-timing PMT (FPMT), has gained significant attention across various fields due to its single-photon detection capability and great time resolution. The after-pulses are spurious pulses occurring after the main pulse and cannot be distinguished from the true signals, which means the occurrence of the after-pulses can contribute to the noise of the FPMT. In this study, the waveforms of the FPMTs were recorded by a high-sampling-rate oscilloscope, and we measured the after-pulse characteristics of different types of FPMTs, including the time distribution and the after-pulse rate. The causes of different after-pulse groups are also analyzed. Additionally, we conducted a comparative analysis of the distinct properties of after-pulses in different types of FPMTs.