The Simulation of Centroid Guided Maximum Likelihood Positioning Algorithm

Bai-yin CAI,Wen-tian CAO,Song-yun LING,Shang-lian BAO
2009-01-01
Abstract:Objective: A key problem in SPECT (Single Photon Emission Computed Tomography) is to work out the location of hit by gamma rays in the crystal, from the output signals of the PMTs (Photomultiplier Tubes) in order to acquire the accurate projection data. Based on the systematic analysis of the centroid weighted (CW) and the maximum likelihood (ML) positioning algorithms for SPECT, a method called centroid guided maximum likelihood (CGML) algorithm was realized by simulation. Methods: CW algorithm is used firstly to calculate event position cursorily, and the positions of the events are accurately determined by ML algorithm between the positioning error regions after non-linear calibration. Results: The position estimated by CGML algorithm can accurately inosculate with the primary position, the positioning error can be controlled in 0.2mm, and the computation time was decreased by 30 times than the general ML algorithm. Conclusion: The searching region is greatly re- duced by using the CW algorithm which can reduce the memory requirement and computation burden for the general ML calculation. In conclusion, this algorithm can be used in the design of the detector.
What problem does this paper attempt to address?