A region segmentation based algorithm for building crystal position lookup table in scintillation detector

Hai Peng Wang,Ming Kai Yun,Shuang Quan Liu,Xin Fan,Xue Xiang Cao,Pei Chai,Bao Ci Shan
DOI: https://doi.org/10.48550/arXiv.1404.6628
2014-04-26
Instrumentation and Detectors
Abstract:In scintillation detector, scintillation crystals are typically made into 2-dimension modular array. The location of incident gamma-ray need be calibrated due to spatial response nonlinearity. Generally, position histograms, the characteristic flood response of scintillation detectors, are used for position calibration. In this paper, a position calibration method based on crystal position lookup table which maps the inaccurate location calculated by Anger logic to the exact hitting crystal position has been proposed, Firstly, position histogram is segmented into disconnected regions. Then crystal marking points are labeled by finding the centroids of regions. Finally, crystal boundaries are determined and crystal position lookup table is generated. The scheme is evaluated by the whole-body PET scanner and breast dedicated SPECT detector developed by Institute of High Energy Physics, Chinese Academy of Sciences. The results demonstrate that the algorithm is accurate, efficient, robust and general purpose.
What problem does this paper attempt to address?