Determination of Residual Benzene Homologues in Chlorphenamine Maleate by Headspace Single-Drop Microextraction-Gc

ZHOU Jian-ke,HAN Kang,ZHANG Qian-li,ZHANG Li
2006-01-01
Abstract:To establish a new method for determination residual organic solvent in medicine.Benzene, toluene and p-xylene in chlorphenamine maleate was determined.Methods:Benzene homologues were concentrat- ed by headspace single-drop microextraction and were determinated by GC.The 5% AT-1200+1.75% Bentone 34 Chromosorb W-AW stainless steel packed column(182 cm×0.32 cm)was used and the column temperature was 80 ℃.The flame ionization detector was used and detector temperature was 190 ℃.The injector port tempera- ture was 190 ℃.Carrier gas was nitrogen with a flow rate of 25 mL·min~(-1).Results:The detection limits of ben- zene,toluene and p-xylene were 1.3×10~(-3),1.3×10~(-3) and 7.8×10~(-4) μg·g~(-1) respectively.The average recov- eries were 97.0%,100.2% and 92.7% respectively.Conclusion:This method used simple apparatus,the equilib- rium time was short,the equilibrium temperature was low and the sensitive was high.The toluene was found in chlorphenamine maleate,but it didn't exceed the standard.