Determination of Residual Solvents in Pharmaceutical Waste Water by Headspace Gas Chromatography

Ming Zhu
2010-01-01
Pharmacy Today
Abstract:Objective To establish a method for the determination of residual solvents in pharmaceutical waste water.Methods Headspace gas chromatography was employed.Chromatographic conditions:SE-54 capillary column(30 m×0.53 mm×5 μm) was used.Column temperature was 40℃ initially and maintained for five minutes;then,heated up to 90℃ at the rate 2℃/min and maintained for zero minute;afterwards,heated up to 160℃ at the rate of 30℃/min and maintained for 2 min.Flame ionization detector(FID) was used and its temperature was 260℃;temperature of the injector was 220℃.Nitrogen was used as the carry gas;the flow rate was 8ml/min;and the distributary ratio was 10∶1.Results The calibration curves for methanol,acetone,n-butyl alcohol,toluene,and butyl acetate were linear in their respective ranges of 0.079~1.582 mg/ml,0.079~1.576 mg/ml,0.081~1.620 mg/ml,0.052~1.038 mg/ml and 0.070~1.410 mg/ml.The average recovery rates(n=9) were between 100.3%~104.1%.Conclusion This method is easy to operate with precise results,and reliable for determining the residual solvents in pharmaceutical wastewater.
What problem does this paper attempt to address?